• 文献标题:   Paper-based amperometric sensor for determination of acetylcholinesterase using screen-printed graphene electrode
  • 文献类型:   Article
  • 作  者:   PANRAKSA Y, SIANGPROH W, KHAMPIENG T, CHAILAPAKUL O, APILUX A
  • 作者关键词:   paperbased sensor, screenprinted graphene electrode, acetylcholinesterase
  • 出版物名称:   TALANTA
  • ISSN:   0039-9140 EI 1873-3573
  • 通讯作者地址:   Chulalongkorn Univ
  • 被引频次:   18
  • DOI:   10.1016/j.talanta.2017.08.096
  • 出版年:   2018

▎ 摘  要

A simple paper-based sensor with electrochemical detection was first developed for rapid, selective, and sensitive determination of acetylcholinesterase (AChE). The screen-printed graphene electrode was used as working electrode providing sensitivity for the sensor. The amperometric detection of AChE is based on the determination of thiocholine (TCh) produced from hydrolysis of acetylthiocholine chloride (ATCh) by AChE. To detect TCh, the ATCh immobilized sheet was stacked onto the detection sheet using double adhesive tape, then samples of AChE were dropped on the back side of an ATCh immobilized sheet with only 1 min of incubation time. To avoid interference, glutathione (GSH), the potential of 0.5 V vs. Ag/AgCl was applied onto a graphene electrode and the current, which depends on AChE concentration, was measured. Under optimized conditions, the limit of detection (LOD) from the experiment of AChE determination was 0.1 U/mL with AChE concentration in range of 0.1-15 U/mL. The data correlated well with the data obtained using spectrophotometric method. The developed sensor had been successfully applied to detect AChE in blood samples.