▎ 摘 要
A novel hexagonal boron nitride modified reduced graphene oxide material was synthesized and used as the adsorbent for the solid-phase microextraction of seven polycyclic aromatic hydrocarbons from water and soil samples prior to their detection by gas chromatography-flame ionization detector. Under optimal conditions, the linear response range of the analytes for water sample is 0.25-50 ng/mL with the correlation coefficients (r) ranging between 0.9953 and 0.9996. The linear range for soil sample is 1.0-400 ng/g with r ranging from 0.9959 to 0.9999. On the basis of the signal-to-noise ratio of 3, the limits of detections for the analytes ranged from 0.05 to 0.15 ng/mL for water samples, and from 0.3 to 0.5 ng/g for soil samples. The relative recoveries of the seven polycyclic aromatic hydrocarbons for water and soil samples were in the range of 79.55-120.0 and 78.76-120.8%, respectively. The relative standard deviations for the determination of the analytes in water and soil samples were lower than 11 and 10%, respectively. The method is simple and suitable for the determination of polycyclic aromatic hydrocarbon residues in water and soil samples.