• 文献标题:   Application of Graphene as a Sorbent for the Preconcentration and Determination of Trace Amounts of Mercury in Water Samples by Hydride Generation Atomic Fluorescence Spectrometry
  • 文献类型:   Article
  • 作  者:   LIU HY, MENG XF, LI C, JIA AQ, LI JC, MA JJ
  • 作者关键词:   graphene, solid phase extraction, preconcentration, hydride generation atomic fluorescence spectrometry, mercury
  • 出版物名称:   JOURNAL OF THE CHEMICAL SOCIETY OF PAKISTAN
  • ISSN:   0253-5106
  • 通讯作者地址:   Agr Univ Hebei
  • 被引频次:   1
  • DOI:  
  • 出版年:   2015

▎ 摘  要

The potential of graphene as a solid-phase extraction adsorbent for the separation and preconcentration of mercury (Hg) was investigated. Hg2+ was found to be quantitatively adsorbed onto graphene within pH 6.0-8.0, and then completely eluted with 4.0 mL of nitric acid/methanol (1:1, v/v) solution at a flow rate of 2.0 mL min(-1). A new method using a microcolumn packed with graphene as a sorbent was developed for the preconcentration of trace amount of Hg2+ prior to its determination by hydride generation atomic fluorescence spectrometry. Under the optimum experimental conditions, the detection limit of this method for Hg2+ was 5.0 ng L-1, with an enrichment factor of 15.0, and the relative standard deviation was 3.5% at the 1.0 mu g L-1 Hg2+ level. The method was then applied for the determination of trace amount of Hg2+ in water samples with satisfactory results.