• 文献标题:   Immobilization-Free Electrochemical Sensor Coupled with a Graphene-Oxide-Based Aptasensor for Glycated Albumin Detection
  • 文献类型:   Article
  • 作  者:   WAIWINYA W, PUTNIN T, PIMALAI D, CHAWJIRAPHAN W, SATHIRAPONGSASUTI N, JAPRUNG D
  • 作者关键词:   graphene oxide, glycated human serum albumin, immobilizationfree electrochemical measurement, screenprinted carbon electrode, diabetes mellitu
  • 出版物名称:   BIOSENSORSBASEL
  • ISSN:  
  • 通讯作者地址:  
  • 被引频次:   15
  • DOI:   10.3390/bios11030085
  • 出版年:   2021

▎ 摘  要

An immobilization-free electrochemical sensor coupled with a graphene oxide (GO)-based aptasensor was developed for glycated human serum albumin (GHSA) detection. The concentration of GHSA was monitored by measuring the electrochemical response of free GO and aptamer-bound GO in the presence of glycated albumin; their currents served as the analytical signals. The electrochemical aptasensor exhibited good performance with a base-10 logarithmic scale. The calibration curve was achieved in the range of 0.01-50 mu g/mL. The limit of detection (LOD) was 8.70 ng/mL. The developed method was considered a one-drop measurement process because a fabrication step and the probe-immobilization process were not required. This simple sensor offers a cost-effective, rapid, and sensitive detection method, and could be an alternative approach for determination of GHSA levels.