• 文献标题:   Development of an unmodified screen-printed graphene electrode for nonenzymatic histamine detection
  • 文献类型:   Article
  • 作  者:   NAKTHONG P, KONDO T, CHAILAPAKUL O, SIANGPROH W
  • 作者关键词:  
  • 出版物名称:   ANALYTICAL METHODS
  • ISSN:   1759-9660 EI 1759-9679
  • 通讯作者地址:   Srinakharinwirot Univ
  • 被引频次:   0
  • DOI:   10.1039/d0ay01443e
  • 出版年:   2020

▎ 摘  要

The high requirement for food quality control has inspired the creation of high-performance sensing, cost-effectiveness, and ease to use. Therefore, the aim of this work is to develop nonenzymatic electrochemical platforms for direct detection of histamine using unmodified screen-printed graphene electrodes (SPGEs) for their applications such as evaluation of fish freshness. In alkaline media (0.2 M NaOH), unmodified SPGEs showed a very low oxidation potential of histamine at +0.58 V (vs. Ag/AgCl) avoiding perturbations from other biogenic amines. The developed method offers an excellent selectivity, sensitivity (a limit of detection (at 3SD/slope) of 0.62 mg L-1) and wide working linear range (5-100 mg L-1) for histamine detection. In addition, the proposed method was successfully applied to detect histamine in canned fish samples with recovery values ranging from 90.72% to 101.21%. Therefore, this newly proposed method is promising as an alternative choice for the determination of histamine in fish samples and related food products.