• 文献标题:   Graphene Nanoplatelets-Based Aptamer Biochip for the Detection of Lipocalin-2
  • 文献类型:   Article
  • 作  者:   MATASSAN ND, RIZWAN M, MOHDNAIM NF, TLILI C, AHMED MU
  • 作者关键词:   electrode, protein, graphene, biosensor, immune system, electrochemical aptasensor, graphene nanoplatelet, lipocalin, screenprinted carbon electrode, voltammetry
  • 出版物名称:   IEEE SENSORS JOURNAL
  • ISSN:   1530-437X EI 1558-1748
  • 通讯作者地址:   Univ Brunei Darussalam
  • 被引频次:   0
  • DOI:   10.1109/JSEN.2019.2927139
  • 出版年:   2019

▎ 摘  要

An electrochemical label-free aptasensor platform was developed by using amine-functionalized aptamer that was incorporated onto screen-printed carbon electrode (SPCE) modified with graphene nanoplatelets (GNPs) for the detection of lipocalin-2 (LCN-2) protein. The aptamers were covalently bound to the GNPs surface via 1-Ethyl-3-(3-dimethylaminopropyl)-carbodiimide (EDC) and N-hydroxysuccinimide (NHS) chemistry and were used as the molecular recognition element for the target protein. Successful interaction between aptamer and LCN-2 protein was ascertained via electrochemical probe, ferro-, and ferri-cyanide {Fe(CN)(6)(3-/4-)} and monitored by means of square wave voltammetry (SWV) whilst layer-by-layer fabrication study was done using cyclic voltammetry (CV). The developed LCN-2 aptasensor achieved sub-picomolar detection of the protein, with the detection limit determined as 0.07 pg ml(-1), and a dynamic linear range of 0.1 to 10 pg ml(-1). Fabricated aptasensor successfully demonstrated high selectivity, stability, and reproducibility as well as displaying potential for the in vitro detection of LCN-2 in real serum samples.