• 文献标题:   Amperometric aptasensor for carcinoembryonic antigen based on a reduced graphene oxide/gold nanoparticles modified electrode
  • 文献类型:   Article
  • 作  者:   VILLALONGA A, VEGAS B, PANIAGUA G, EGUILAZ M, MAYOL B, PARRADO C, RIVAS G, DIEZ P, VILLALONGA R
  • 作者关键词:   biosensor, carcinoembryonic antigen, graphene, gold nanoparticle, aptamer
  • 出版物名称:   JOURNAL OF ELECTROANALYTICAL CHEMISTRY
  • ISSN:   1572-6657 EI 1873-2569
  • 通讯作者地址:   Univ Complutense Madrid
  • 被引频次:   0
  • DOI:   10.1016/j.jelechem.2020.114511
  • 出版年:   2020

▎ 摘  要

We report the construction of a novel amperometric aptasensor for the highly specific detection of carcinoembryonic antigen (CEA). The sensing interface was assembled by electrodeposition of reduced graphene oxide (rGO) and gold nanoparticles (AuNPs) on a carbon screen-printed electrode (SPE) and further functionalization with a biotin and thiol-modified anti-CEA DNA hairpin aptamer. The sensing approach relies on the specific recognition of CEA by the folded aptamer, causing unfolding of the DNA hairpin structure and unmasking the biotin residues at the aptamer chain. Further incubation with a sptreptavidin-peroxidase conjugate allows the ampemmetric detection of the cancer biomarker. This aptasensor was able to detect CFA in the broad range from 20 pg mL(-1) to 2 mu g mL(-1) (112 fM-11 mu M) with a detection limit of 16 pg mL(-1) (90 fm). The aptasensor also showed high reproducibility, specificity and stability, and was successfully validated in diluted human serum samples.