• 文献标题:   Application of Graphene as a Sorbent for Preconcentration and Determination of Trace Amounts of Cadmium
  • 文献类型:   Article
  • 作  者:   WANG YK, DONG YL, ZHANG JW, LI C, LI JC, MA JJ
  • 作者关键词:  
  • 出版物名称:   CHEMISTRY LETTERS
  • ISSN:   0366-7022 EI 1348-0715
  • 通讯作者地址:   Agr Univ Hebei
  • 被引频次:   3
  • DOI:   10.1246/cl.2013.153
  • 出版年:   2013

▎ 摘  要

A new method using a column packed with graphene as sorbent was developed for the preconcentration of trace amounts of cadmium (Cd) prior to its determination by flame atomic absorption spectrometry. Under the optimal conditions, the calibration graph was linear in the concentration range of 10.0-150.0 mu g L-1 with a detection limit of 0.60 mu g L-1. The relative standard deviation for ten replicate measurements of 100.0 mu g L-1 of Cd was 3.25%.