• 文献标题:   A sensitive electrochemical platform integrated with a 3D graphene aerogel for point-of-care testing for tumor markers
  • 文献类型:   Article
  • 作  者:   HU T, WU ZH, SANG W, DING B, CHEN K, LI X, SHEN Y, NI ZH
  • 作者关键词:  
  • 出版物名称:   JOURNAL OF MATERIALS CHEMISTRY B
  • ISSN:   2050-750X EI 2050-7518
  • 通讯作者地址:  
  • 被引频次:   2
  • DOI:   10.1039/d2tb00846g EA AUG 2022
  • 出版年:   2022

▎ 摘  要

Point-of-care testing (POCT) of tumor markers, such as alpha-fetoprotein (AFP) and carcinoembryonic antigen (CEA), can be used for the early diagnosis of cancer. In this paper, a highly sensitive electrochemical immuno-biochip based on a porous three-dimensional graphene aerogel (3D-GA) is presented to detect multiple tumor biomarkers and exosomes. The 3D-GA was prepared via in situ chemical reduction of graphene oxide with l-ascorbic acid and then dehydration by freeze-drying. The obtained 3D-GA exhibits a large specific surface area of 125.3 m(2) g(-1) due to its intrinsic 3D porous architecture. After chemical activation and modification of the 3D-GA, the prepared microfluidic biochip can be used for detecting various tumor markers in liquid samples via electrochemical impedance spectroscopy (EIS). The electrochemical platform with only 5 mu L sample achieved a broad detection range of 1.0 x 10(-8)-1.0 x 10(-5) and 1.0 x 10(-8)-5.0 x 10(-4) mg mL(-1) for AFP and CEA, respectively, and a low limit of detection (LOD) of 7.9 and 6.2 pg mL(-1) for AFP and CEA respectively, which was much better than the outcomes of many other reports. Moreover, the biochip determined the tumor cell-derived exosomes with a low LOD of 10 particles per mu L in the PBS solution and an average recovery rate of similar to 90% in the diluted serum.