• 文献标题:   Simultaneous voltammetric determination of dopamine and uric acid using carbon-encapsulated hollow Fe3O4 nanoparticles anchored to an electrode modified with nanosheets of reduced graphene oxide
  • 文献类型:   Article
  • 作  者:   SONG H, XUE GP, ZHANG JJ, WANG G, YE BC, SUN SG, TIAN LP, LI YC
  • 作者关键词:   electrochemical sensor, glassy carbon electrode, differential pulse voltammetry, graphene oxide nanosheet, whole blood analysi, transmission electron microscopy, scanning electron microscopy
  • 出版物名称:   MICROCHIMICA ACTA
  • ISSN:   0026-3672 EI 1436-5073
  • 通讯作者地址:   Shihezi Univ
  • 被引频次:   2
  • DOI:   10.1007/s00604-016-2067-1
  • 出版年:   2017

▎ 摘  要

The authors describe an electrochemical sensor for simultaneous determination of dopamine (DA) and uric acid (UA). The sensor is based on the use of composite materials of carbon-encapsulated hollow magnetite (Fe3O4) nanoparticles and graphene oxide nanosheets, which were homogeneously deposited on a glassy carbon electrode. The morphology of the materials was characterized by transmission electron microscopy and scanning electron microscopy, and the properties of differently decorated electrodes were investigated via cyclic voltammetry and differential pulse voltammetry (DPV). Simultaneous detection of DA and UA was carried out via DPV, and the potentials of electrooxidation peaks are 0.14 V and 0.27 V (vs. SCE) for DA and UA, respectively. The calibration plot is linear from 0.1 to 150 mu M for DA, and from 1.0 to 100 mu M for UA, with detection limits of 0.053 mu M and 0.41 mu M at an S/N ratio of 3. The method was applied to the quantification of DA and UA in brain tissue, urine and whole blood.