• 文献标题:   An electrochemical immunosensor for simultaneous point-of-care cancer markers based on the host-guest inclusion of beta-cyclodextrin-graphene oxide
  • 文献类型:   Article
  • 作  者:   FENG TT, QIAO XW, WANG HN, SUN Z, QI Y, HONG CL
  • 作者关键词:  
  • 出版物名称:   JOURNAL OF MATERIALS CHEMISTRY B
  • ISSN:   2050-750X EI 2050-7518
  • 通讯作者地址:   Shihezi Univ
  • 被引频次:   25
  • DOI:   10.1039/c5tb02480c
  • 出版年:   2016

▎ 摘  要

A novel electrochemical immunosensor was developed for the simultaneous detection of alpha-fetoprotein (AFP) and carcinoembryonic antigen (CEA) using Cu2O-graphene oxide-beta-cyclodextrin (Cu2O-GO-CD) and beta-cyclodextrin-graphene oxide-ferrocenecarboxylic acid (GO-CD-Fc-COOH) as the distinguishable signal probe, and graphene oxide-gold nanoparticles (GO-AuNPs) as the sensor platform. GO-CD displayed excellent solubility in water and rich capture capability to Fc-COOH and the secondary antibody. The proposed immunosensor exhibited an excellent electrochemical performance. The linear ranges were from 0.001 ng mL(-1) to 80 ng mL(-1) for AFP and CEA with a detection limit of 0.0002 ng mL(-1) for AFP and 0.0001 ng mL(-1) for CEA. With the merits of acceptable stability, high sensitivity, a wide linear range and a low detection limit, the proposed immunosensor showed great potential for the simultaneous detection of multi-analytes in clinical diagnosis.