▎ 摘 要
A sensitive electrochemical sensor was fabricated based on ceria-graphene oxide nanoribbons composite (CeO2-GONRs) for an antiviral drug, entecavir (ETV). It was characterized by SEM, EDAX, AFM, IR and Raman spectroscopic techniques. The electrochemical behaviour of ETV was investigated by cyclic voltammetric, differential pulse voltammetric (DPV), linear sweep voltammetric (LSV) and square wave voltammetric (SWV) methods at CeO2-GONRs modified glassy carbon electrode. Good linearity was observed between the peak current and concentration of ETV in the range of 0.51 - 100M with a detection limit of 0.042 M in DPV method, 2.1 - 61.1M with a detection limit of 0.7M in LSV method and 0.1 - 80M with a detection limit of 68.1 nM in SWV method. The proposed sensitive DPV method was successfully applied for the determination ETV in tablets and biological samples.