▎ 摘 要
On the surface of a glassy carbon electrode, electrochemically reduced graphene oxide-cationic pillar[6]arene (ErGO-CP6) composite film was constructed using a pulsed potential method. UV-vis spectra, SEM, Raman spectra and electrochemical experiments were applied to characterize the composite film. It was then used as a new electrochemical sensing platform for determination of thiamethoxam. Due to the synergistic effect of ErGO and CP6, this composite film shows a higher sensitivity and better selectivity toward thiamethoxam than that of ErGO film. The linear range from 1.0 x 10(-7) to 1.3 x 10(-5) mol l(-1) was obtained by differential pulse voltammetry. Meanwhile, the method was applied to cucumber and tomato samples in a recovery test. The recovery was between 92.0% and 98.7%, and the results are satisfactory. This study presents a promising electrochemical sensing platform for rapid and sensitive analysis of thiamethoxam.