• 文献标题:   A copper oxide-ionic liquid/reduced graphene oxide composite sensor enabled by digital dispensing: Non-enzymatic paper-based microfluidic determination of creatinine in human blood serum
  • 文献类型:   Article
  • 作  者:   BOOBPHAHOM S, RUECHA N, RODTHONGKUM N, CHAILAPAKUL O, REMCHO VT
  • 作者关键词:   paperbased analytical device, nonenzymatic sensor, creatinine, copper oxide, ionic liquid, electrochemically reduced graphene oxide, hp d300 digital dispenser
  • 出版物名称:   ANALYTICA CHIMICA ACTA
  • ISSN:   0003-2670 EI 1873-4324
  • 通讯作者地址:   Chulalongkorn Univ
  • 被引频次:   11
  • DOI:   10.1016/j.aca.2019.07.029
  • 出版年:   2019

▎ 摘  要

A paper-based analytical device (PAD) with an integrated composite electrode for non-enzymatic creatinine sensing was developed. The electrode was produced and optimization was efficiently accomplished using a rapid digital dispensing approach. The electrochemical sensor was fabricated using an HP D300 digital dispenser to deliver a copper oxide and ionic liquid composite onto an electrochemically reduced graphene modified screen-printed carbon electrode (CuO/IL/ERGO/SPCE) on a PAD. The modified electrode was characterized using electrochemical and microscopic techniques. Electrochemical detection of creatinine was performed on the SPCE using amperometry at a constant potential. Under optimized conditions, the paper-based sensor exhibited a linear range of 0.01-2.0 mM (R-2 = 0.99) and the limit of detection was 0.22 mu M (S/N = 3, IUPAC definition) for creatinine. The simple fabrication process, low cost, and clinically appropriate creatinine sensitivity make this device applicable for pointof-care use. (C) 2019 Elsevier B.V. All rights reserved.