• 文献标题:   Solid-contact polymeric membrane ion-selective electrodes using a covalent organic framework@reduced graphene oxide composite as ion-to-electron transducer
  • 文献类型:   Article
  • 作  者:   ZHANG WT, LI JH, QIN W
  • 作者关键词:   covalent organic framework, solid contact, ionselective electrode, reproducibility, stability
  • 出版物名称:   TALANTA
  • ISSN:   0039-9140 EI 1873-3573
  • 通讯作者地址:  
  • 被引频次:   0
  • DOI:   10.1016/j.talanta.2023.124444 EA MAR 2023
  • 出版年:   2023

▎ 摘  要

A solid-contact ion-selective electrode (SC-ISE) based on a covalent organic framework@reduced graphene oxide (rGO) composite is proposed. The composite can be synthesized through the polycondensation of 1,3,5-triformyl-phloroglucinol (TFP) and 2,6-diaminoanthraquinone (DAAQ) on the rGO nanosheets, which shows high capacitance and good redox-active properties. By applying Cd2+-ISE as a model, the electrode exhibits a Nernstian slope of 29.7 +/- 0.4 mV/decade in the activity range of 1.0 x 10-7 -7.9 x 10-4 M and the limit of detection is 6.8 x 10-8 M. Particularly, the electrode based on DAAQ-TFP@rGO exhibits a low potential drift of 1.2 +/- 0.2 mu V/h over 70 h due to the large capacitance of 2.0 mF. Moreover, the DAAQ-TFP@rGO-based Cd2+-ISE shows good reproducibility and the standard deviations of the standard potentials for single batch and batch-to-batch are 0.28 (n = 4) and 0.30 mV (n = 4), respectively. The developed SC-Cd2+-ISE is not disturbed by light or gas and no aqueous layer occurs between the sensing membrane and DAAQ-TFP@rGO layer. The DAAQ-TFP@rGO composite is highly promising for construction of calibration-free SC-ISEs.