• 文献标题:   All Solid State Graphene Based Potentiometric Sensors for Monitoring of Mercury Ions in Waste Water Samples
  • 文献类型:   Article
  • 作  者:   GANJALI MR, FARIDBOD F, DAVARKHAH N, SHAHTAHERI SJ, NOROUZI P
  • 作者关键词:   all solid state, mercury, potentiometry, sensor
  • 出版物名称:   INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH
  • ISSN:   1735-6865 EI 2008-2304
  • 通讯作者地址:   Univ Tehran
  • 被引频次:   31
  • DOI:  
  • 出版年:   2015

▎ 摘  要

In this work, all-solid-state potentiometric PVC membrane sensor was introduced for fast and simple monitoring of mercury ions in waste water samples. The all solid sate electrode is made based on a conductive composite of graphite, reduced graphene oxide, epoxy resin on a copper wire. A thin layer PVC membrane is then coated on the surface of the new conducting transducer. PVC membrane is composed of 30% PVC, 62% NPOE as plasticizer, 3% ionic liquid, and 5% He2+ ion imprinted polymer (IIP). The sensing element of the sensor is an artificial host which is imprinted by vinyl pyridine and a suitable cross-linker. The sensor shows a Nernstian response of 29.4 +/- 0.3 mV/decade. Dynamic linear range of the sensor is 1.0 x 10(-8)-1.0 x 10(-3) mol/L of the mercury concentration and detection limit of 3.2 x 10(-9) mol/L. The interference of different ionic species with the response of the electrode shows a good selectivity of the proposed sensor. The proposed sensor is successfully applied in determination of mercury ions in some waste water samples.