• 文献标题:   Electroanalytical method for the determination of 5-fluorouracil using a reduced graphene oxide/chitosan modified sensor
  • 文献类型:   Article
  • 作  者:   PATTAR VP, NANDIBEWOOR ST
  • 作者关键词:  
  • 出版物名称:   RSC ADVANCES
  • ISSN:   2046-2069
  • 通讯作者地址:   Karnatak Univ
  • 被引频次:   16
  • DOI:   10.1039/c5ra04396d
  • 出版年:   2015

▎ 摘  要

A glassy carbon electrode (GCE) modified with a chemically reduced graphene oxide and chitosan (CRGO/CS) composite film was constructed and used to determine 5-fluorouracil using cyclic, staircase and square wave voltammetric techniques. The synthesized graphene oxide material was characterized by XRD, FTIR and UV spectroscopic techniques. The surface morphology of the CRGO/CS film was characterized using atomic force microscopy. The oxidation potential shifted negatively compared to GCE, indicating that the CRGO/CS modified electrode had obvious catalytic activity for the 5FU oxidation. The effects of pH, scan rate and excipients were studied. The sensor showed a good detection limit and it was successfully applied to determine 5FU in pharmaceutical and human biological samples.