▎ 摘 要
In this study, a novel sensitive electrochemiluminescence (ECL) immunosensor was constructed by carboxyl graphene (GR) for enhancing luminol-O-2 system emission. Here, carboxyl GR was used to enhance the ECL intensity of luminol that had excellent electron transfer ability and good solubility. The sensing platform was constructed by depositing carboxyl GR on electrodes and immobilizing antibodies on the surface of carboxyl GR through amidation. The specific immunoreaction between alpha-fetoprotein (AFP) and antibodies resulted in a decrease of ECL intensity, and the intensity decreased linearly with AFP concentrations in the range of 5 pg ml(-1) to 14 ng ml(-1) with a detection limit of 2.0 pg ml(-1). The proposed immunosensor exhibits high specificity, good reproducibility, and longtime stability. It may become a promising technique for protein detection. (C) 2014 Elsevier Inc. All rights reserved.