• 文献标题:   A directional fibre optic ultrasound transmitter based on a reduced graphene oxide and polydimethylsiloxane composite
  • 文献类型:   Article
  • 作  者:   COLCHESTER RJ, ALLES EJ, DESJARDINS AE
  • 作者关键词:  
  • 出版物名称:   APPLIED PHYSICS LETTERS
  • ISSN:   0003-6951 EI 1077-3118
  • 通讯作者地址:   UCL
  • 被引频次:   5
  • DOI:   10.1063/1.5089750
  • 出版年:   2019

▎ 摘  要

Strongly directional ultrasound sources are desirable for many minimally invasive applications, as they enable high-quality imaging in the presence of positioning uncertainty. All-optical ultrasound is an emerging paradigm that exhibits high frequencies, large bandwidths, and a strong miniaturisation potential. Here, we report the design, modelling, and fabrication of a highly directional fibre-optic ultrasound transmitter that uses a composite of reduced graphene oxide and polydimethylsiloxane as the optical ultrasound generator. The ultrasound transmitter, which had an outer diameter of just 630 lm, generated ultrasound with a pressure exceeding 0.4 MPa for axial distances up to 16 mm, at a large bandwidth of 24.3 MHz. The ultrasound beam exhibited low divergence, with a beam diameter ranging between 0.6 and 2.1mm for distances between 0 and 16 mm. The presented directional optical ultrasound source is hence well-suited to high-resolution interventional imaging. (C) 2019 Author(s).