▎ 摘 要
The studies considered graphene-based biomaterials dedicated for cardiovascular therapy. Reduced graphene oxide flakes were introduced into the porous structure of the polyelectrolyte based coatings. TEM analysis showed the presence of graphene flakes arranged parallel to the substrate surface, firmly connected to the porous coating. Biomaterials were subjected to a comprehensive diagnosis of the biological and material properties. The material behavior was simulated using finite element method. The coatings were deposited using layer by layer method. Mechanical analysis was done using Berkovich indenter geometry. They confirmed theoretical PEA based calculations, it was observed the coating stiffness incensement under the influence of introduced particles of graphene. The endanger of the bacteriology film formation was verified based on the E-coli and Staphylo coccus bacteria. Blood material interaction was examined in the dynamic flow conditions. Bacteriological analysis showed reduced presence of bacteria after contact with the surface with introduced graphene flakes. Dynamic analyzes on blood showed high activation of the coagulation, strong platelets consumption and a strong immune response. It is caused by sharp edge of the single plane of the graphene flake. (C) 2014 Politechnika Wroclawska. Published by Elsevier Urban & Partner Sp. z o.o. All rights reserved.