• 文献标题:   Electrochemical Characterization of Immobilized Glucose Oxidase on Matrix of Graphene-Ion Liquid Composite
  • 文献类型:   Article
  • 作  者:   LIU HY, LIU Y, ZHANG XY, XU ZM, LIU BA
  • 作者关键词:   graphene, ionic liquid, glucose oxidase, direct electron transfer, biosensor
  • 出版物名称:   ASIAN JOURNAL OF CHEMISTRY
  • ISSN:   0970-7077
  • 通讯作者地址:   Jilin Univ
  • 被引频次:   0
  • DOI:  
  • 出版年:   2012

▎ 摘  要

The graphene-ion liquid nanocomposite (IL-GNs), characterized by good conductivity, high stability and good biocompatibility, provides a suitable biosensing matrix. Enzymes can be firmly incorporated into the matrix without the aid of other cross-linking reagents. Based on this, the g ucose oxidase (GOD) is entrapped in the composite of IL-GNs and the direct electron transfer reaction between glucose oxidase and electrode takes place. The electron transfer rate of glucose oxidase is greatly enhanced to 6.7 s(-1) in the system. It demonstrates that the conformational change of glucose oxidase in the microenvironment enables the accessibility of active site for glucose oxidase to the electrode. In addition, a third-generation glucose biosensor has been developed. The amperometric response of this sensor to glucose is observed with a linear range between 0.2 rnM to 5 mM and a detection limit of 30 mu M. The facile procedure of immobsilizing glucose oxidase will promote the developments of electrochemical research for protein, biosensors and other bio-electrochemical devices.