• 文献标题:   Sensitive Quantification of Fe(III) in Food Samples at Screen Printed Carbon Electrode Modified with Graphene and Piroxicam by Catalytic Adsorptive Voltammetry
  • 文献类型:   Article
  • 作  者:   SADEGHI S, EBADI S
  • 作者关键词:   fe iii, screen printed carbon electrode, graphene, piroxicam, catalytic adsorptive voltammetry
  • 出版物名称:   ELECTROANALYSIS
  • ISSN:   1040-0397 EI 1521-4109
  • 通讯作者地址:   Univ Birjand
  • 被引频次:   0
  • DOI:   10.1002/elan.202060068 EA JUN 2020
  • 出版年:   2020

▎ 摘  要

A new disposable sensitive voltammetric sensor for the determination of Fe(III) based on a graphene (G) and piroxicam (Pir) modified screen printed carbon electrode (Pir/G/SPCE) has been developed. The developed method is based on accumulation of Fe(III) on the surface of the prepared sensor strip, formation a complex with Pir and subsequent reduction the adsorbed chelated Fe(III) at -0.03 V (vs. Ag/AgCl) coupled with the catalytic enhancement of bromate. Characterizations of the modified electrode surface were performed by field emission scanning electron microscopy (FE-SEM), energy dispersive X-ray spectroscopy (EDX) and electrochemical impedance spectroscopy (EIS). Electrochemical behavior of the modified SPCEs was investigated by cyclic voltammetry (CV) and differential pulse voltammetry (DPV). Under the optimum conditions, the catalytic voltammetric method exhibited linear calibration plot in the concentration ranges of 1-100 ng mL(-1) and 100-3500 ng mL(-1) Fe(III) with a limit of detection of 0.3 ng mL(-1). The sensor strip displayed good reproducibility with 1.7 % relative standard deviation (RSD%). The developed method was successfully applied for the determination of iron in food samples such as vegetables, fruit, and cereal.