▎ 摘 要
Determination of Human Serum Albumin (HSA), an important biomarker, is needed for early diagnosis and therapy monitoring in a variety of diseases. A novel electrochemical immunosensor was constructed to detect HSA in both standard solutions and human scrum samples. In so doing, glassy carbon electrode (GCE) was modified in three steps. First, graphene oxide (GO) was drop-tasted on the GCE surface and electmchemically reduced. Second, in situ electrochemical deposition (ECD) of cupric oxide nanoparticles (0.10 NPs) in two consecutive steps has been done. Finally, anti! ISA antibodies have been immobilized via amino groups of chitosan which drop-tasted earlier. The surface morphology and nature of the modified electrode was characterized using common electrochemical and imaging techniques. Using differential pulse voltammetry (DPV), the response of the immunosensor was linear in the range of 10-450 ng.mL(-1) HSA, with the limit of detection (LOD) 2.6 ng.mL(-1). Moreover, the selectivity of the modified electrode was investigated to some common interferences. Under optimum conditions, the immunosensor displayed excellent stability, reproducibility, selectivity and repeatability. The findings indicated that fabricated electrode is a promising candidate for quantifying HSA that is applicable in diagnostics and therapeutics.