▎ 摘 要
Highly dispersed titanium dioxide nanocluster (TDN) was synthesized on reduced graphene oxide (RGO) in a toluene-water system under microwave irradiation. The prepared RGO-TDN hybrids were used to modify glassy carbon electrode for loading glucose oxidase. The fabricated glucose biosensor exhibits excellent performance for glucose sensing including low work potential (-0.7 V), high sensitivity (35.8 mu A mM(-1) cm(-2)), low detection limit (4.8 mu M), wide linear range from 0.032 to 1.67 mM, small Michaelis-Menten constant (K-m) (0.81 mM), and short response time (10 s). (C) 2013 Elsevier Ltd. All rights reserved.