▎ 摘 要
Graphene sheets are used for the first time to fabricate a new type of solid-contact ion-selective electrode (SC-ISE) as the intermediate layer between an ionophore-doped solvent polymeric membrane and a glassy carbon electrode. The new transducing layer was characterized by transmission electron microscopy, scanning electron microscopy, cyclic voltammetry and electrochemical impedance spectroscopy. The performance of the new K(+-)selective electrodes was examined by a potentiometric water layer test, potentiometric measurements, and current reversal chronopotentiometry. The obtained potentiometric sensors were characterized with a calibration line of slope close to Nernstian (59.2 mV/decade) within the activity from 10(-4.5) to 0.1 M. The high capacitance of the graphene solid contacts results in a signal that is stable over one week. The short response time is less than 10 s for activities higher than 10(-5) M. The potential drift of the electrodes was calculated from the slope of the curves at longer times (Delta E/Delta t = 1.2 x 10(-5) V s(-1) (I = 1 nA) and Delta E/Delta t = 5.5 x 10(-5) V s(-1) (I = 5 nA)). All the results indicate that graphene is a promising material for use as a transducer layer for SC-ISEs.