• 文献标题:   Determination of Tedizolid Phosphate Using Graphene Nanocomposite Based Solid Contact Ion Selective Electrode; Green Profile Assessment by Eco-scale and GAPI Approach
  • 文献类型:   Article
  • 作  者:   MOAAZ EM, MAHMOUD AM, FAYED AS, REZK MR, ABDELMOETY EM
  • 作者关键词:   tedizolid phosphate, graphene nanocomposite, solidcontact ion selective electrode, glassy carbon electrode, green chemistry
  • 出版物名称:   ELECTROANALYSIS
  • ISSN:   1040-0397 EI 1521-4109
  • 通讯作者地址:  
  • 被引频次:   11
  • DOI:   10.1002/elan.202100067 EA MAY 2021
  • 出版年:   2021

▎ 摘  要

Antibiotics determination plays a major role in minimizing antimicrobial resistance starting from quality control of pharmaceutical formulations to therapeutic drug monitoring. Green modified glassy carbon electrode has been developed for determination of tedizolid phosphate; new antibiotic prodrug; in presence of its active metabolite. The graphene transducer interlayer, dispersed with PVC, improved the electrode stability and standard potential reproducibility. Graphene hydrophobicity prevented the water layer formation between the sensing layers that decreased the potential drift down to 267 mu V h(-1). Electrochemical impedance showed a low resistance value for graphene containing sensor due to its high electron transfer ability.