• 文献标题:   Voltammetric Determination of Bergenin with Graphene Modified Glassy Carbon Electrode
  • 文献类型:   Article
  • 作  者:   LI XB, ZOU RY, NIU YY, SHAO TM, CHEN Y, SUN W, HE MX
  • 作者关键词:   graphene, bergenin, differential pulse voltammetry, electroanalysi
  • 出版物名称:   INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE
  • ISSN:   1452-3981
  • 通讯作者地址:   Hainan Normal Univ
  • 被引频次:   2
  • DOI:   10.20964/2018.02.68
  • 出版年:   2018

▎ 摘  要

This study utilized a sensitive electroanalytical method for the detection of bergenin in phosphate buffer solution (pH 5.0) with a graphene (GR) and chitosan (CS) nanocomposite modified glassy carbon electrode (GCE). Characterization of CS-GR/GCE was carried out by electrochemical impedance spectroscopy. Electrocatalytic oxidation of bergenin on CS-GR/GCE was greatly enhanced with the improvement of the anodic peak current, which allowed the development of a voltammetric sensor for bergenin determination. Under the optimized conditions with differential pulse voltammetry, the anodic peak responses enhanced linearly with bergenin concentration from 2.0x10(-8) mol/L to 8.0x10(-6) mol/L with the detection limit of 3.6 nmol/L (3 sigma). This electrochemical sensor showed high sensitivity and selectivity, excellent reproducibility and good stability. The analytical application of this sensor was successfully proved by the Tabellae Bergenini Compositae sample detection.