• 文献标题:   Electrochemically prepared three-dimensional reduced graphene oxide-polyaniline nanocomposite as a solid-phase microextraction coating for ethion determination
  • 文献类型:   Article
  • 作  者:   HEYDARI M, SARAJI M, JAFARI MT
  • 作者关键词:   negative corona dischargeion mobility spectrometry, solidphase microextraction, threedimensional reduced graphene oxide, polyaniline, ethion
  • 出版物名称:   TALANTA
  • ISSN:   0039-9140 EI 1873-3573
  • 通讯作者地址:   Isfahan Univ Technol
  • 被引频次:   2
  • DOI:   10.1016/j.talanta.2019.120576
  • 出版年:   2020

▎ 摘  要

Here, a three-dimensional reduced graphene oxide-polyaniline nanocomposite was synthesized and applied as a sorbent for solid-phase microextraction of ethion from water samples followed by negative corona discharge ionization ion mobility spectrometry. The nanocomposite was synthesized via a fast and facile two-step electrochemical method. The prepared sorbent was characterized by Fourier transform infrared spectroscopy, field emission scanning electron microscopy and thermogravimetric analysis. The main experimental parameters in the preparation of the fiber including concentration of graphene oxide, electrodeposition time, aniline concentration, and electropolymerization time were investigated. Also, parameters affecting the extraction process (i.e., salt addition, sample stirring rate, solution pH, extraction temperature and extraction time) were studied. Under the optimal conditions, the linearity and limit of detection were found to be 1.0-70 and 0.4 mu g L-1, respectively. The method presented good precision (intra-day and inter-day, n = 3), ranged from 1.7 to 4.7%. The analyte recovery obtained using spiked real water samples were in the range of 84-98%.