• 文献标题:   Capturing hemoglobin on graphene sheet from sub-microliter whole blood for quantitative characterization by internal extractive electrospray ionization mass spectrometry
  • 文献类型:   Article
  • 作  者:   SONG LL, CHEN HW, MA MY, LIU ZW, JIE FF, LAN LQ, HUA R, WANG SL
  • 作者关键词:   quantitative blood sampler, graphene, internal extractive electrospray ionization, mass spectrometry, hemoglobin
  • 出版物名称:   TALANTA
  • ISSN:   0039-9140 EI 1873-3573
  • 通讯作者地址:   East China Univ Technol
  • 被引频次:   1
  • DOI:   10.1016/j.talanta.2019.04.069
  • 出版年:   2019

▎ 摘  要

A disposable blood sampler, which is consisted of a sub-microliter whole blood collector and a graphene filter, loading graphene sheet to selectively capture hemoglobin from sub-microliter whole blood, was developed for both qualitative and quantitative characterization hemoglobin by internal extractive electrospray ionization mass spectrometry (iEESI-MS). The blood collector was elegantly fabricated in syringe-like fashion for precisely sampling tiny amounts (1.0 mu L - 2%) of whole blood, which was immediately diluted by water inside the syringe and was then pressed through the graphene filter placed between the waste outlet and the syringe reservoir to capture the hemoglobin in the blood sample. Then the graphene with hemoglobin was directly eluted by a charged (+2.5 kV) solution (mathanol/water/formic acid, 48/48/4, v/v/v) to produce the hemoglobin ions for mass spectrometric analysis. Low detection-of-limit (19.3 mg L-1 (89.5 picomol)), acceptable linear response range (300-1500 mg L-1, R- (2)= 0.998), relative standard deviation (0.5-6.5%, n = 3), low sample consumption (