• 文献标题:   Amperometric cholesterol biosensor based on zinc oxide films on a silver nanowire-graphene oxide modified electrode
  • 文献类型:   Article
  • 作  者:   WU Q, HOU YT, ZHANG MD, HOU XC, XU L, WANG NN, WANG JP, HUANG W
  • 作者关键词:  
  • 出版物名称:   ANALYTICAL METHODS
  • ISSN:   1759-9660 EI 1759-9679
  • 通讯作者地址:   Nanjing Tech Univ NanjingTech
  • 被引频次:   12
  • DOI:   10.1039/c6ay00158k
  • 出版年:   2016

▎ 摘  要

A highly sensitive and selective cholesterol biosensor is successfully fabricated in this work. Due to excellent conductivity and catalytic activity, silver nanowires (AgNWs) and zinc oxide (ZnO) are modified on an indium tin oxide (ITO) electrode with graphene oxide (GO) and chitosan (CS) to construct the novel amperometric biosensor. The composite of AgNW-ZnO successfully enhanced electron communication, and the electrochemical performances of the ChOx/ZnO/Ag/GO-CS/ITO electrode were investigated via scanning electron microscopy (SEM), cyclic voltammetry (CV) and impedance measurements. The linear response to cholesterol in the range of the fabricated biosensor is 0.25-400 mg dL(-1) (6.5 mu M to 10 mM) with a sensitivity of 9.2 mu A mu M-1 cm(-2). Moreover, the limit of detection (LOD) is 0.287 mu M (S/N = 3), and the Michaelis-Menten constant is calculated to be about 0.295 mu M in 0.1 M phosphate buffer solution (pH 7).