• 文献标题:   Microwave plasma torch-atomic emission spectrometry for the on-line determination of rare earth elements based on flow injection preconcentration by TiO2-graphene composite
  • 文献类型:   Article
  • 作  者:   ZHANG JL, CHENG RM, TONG SS, GU XW, QUAN XJ, LIU YL, JIA Q, JIA JB
  • 作者关键词:   tio2graphene composite, preconcentration, flow injection, rare earth element, microwave plasma torchatomic emission spectrometry
  • 出版物名称:   TALANTA
  • ISSN:   0039-9140
  • 通讯作者地址:   Jilin Univ
  • 被引频次:   23
  • DOI:   10.1016/j.talanta.2011.08.030
  • 出版年:   2011

▎ 摘  要

In this work, we synthesized TiO2-graphene composite as a novel preconcentration material. It was enclosed in a microcolumn in the on-line flow injection system to adsorb trace light (La), medium (Tb), and heavy (Ho) rare earth elements (REEs) prior to their determinations by microwave plasma torch-atomic emission spectrometry (MPT-AES). Various experimental parameters, such as sample loading time, sample flow rate, sample pH, eluent flow rate, eluent concentration, and interfering ions, were investigated systematically. Under the optimum conditions, the detection limits (three times of standard deviations of blank by 7 reiterations) of La, Tb. and Ho were found to be 2.2, 1.6, and 2.8 mu g L-1, with enrichment factors of 17.1, 11.1, and 10.2, respectively. Relative standard deviations for the determination of the target REEs were 3.6%, 1.3%, and 1.4%, respectively (n = 7). The developed method was validated by the analysis of La, Tb, and Ho in certified reference material (GBW07313, marine sediment) and high purity REE oxide samples. (C) 2011 Elsevier B.V. All rights reserved.