• 文献标题:   Metallo-Graphene Nanocomposite Electrocatalytic Platform for the Determination of Toxic Metal Ions
  • 文献类型:   Article
  • 作  者:   WILLEMSE CM, TLHOMELANG K, JAHED N, BAKER PG, IWUOHA EI
  • 作者关键词:   nafiongraphene nanocomposite, mercury film, trace metal, squarewave anodic stripping voltammetry
  • 出版物名称:   SENSORS
  • ISSN:   1424-8220
  • 通讯作者地址:   Univ Western Cape
  • 被引频次:   103
  • DOI:   10.3390/s110403970
  • 出版年:   2011

▎ 摘  要

A Nafion-Graphene (Nafion-G) nanocomposite solution in combination with an in situ plated mercury film electrode was used as a highly sensitive electrochemical platform for the determination of Zn2+, Cd2+, Pb2+ and Cu2+ in 0.1 M acetate buffer (pH 4.6) by square-wave anodic stripping voltammetry (SWASV). Various operational parameters such as deposition potential, deposition time and electrode rotation speed were optimized. The Nafion-G nanocomposite sensing platform exhibited improved sensitivity for metal ion detection, in addition to well defined, reproducible and sharp stripping signals. The linear calibration curves ranged from 1 mu g L-1 to 7 mu g L-1 for individual analysis. The detection limits (3 sigma blank/slope) obtained were 0.07 mu g L-1 for Pb2+, Zn2+ and Cu2+ and 0.08 mu g L-1 for Cd2+ at a deposition time of 120 s. For practical applications recovery studies was done by spiking test samples with known concentrations and comparing the results with inductively coupled plasma mass spectrometry (ICP-MS) analyses. This was followed by real sample analysis.