• 文献标题:   Gold Nanoparticles dotted Reduction Graphene Oxide Nanocomposite Based Electrochemical Aptasensor for Selective, Rapid, Sensitive and Congener-Specific PCB77 Detection
  • 文献类型:   Article
  • 作  者:   WU LD, LU XB, FU XC, WU LX, LIU H
  • 作者关键词:  
  • 出版物名称:   SCIENTIFIC REPORTS
  • ISSN:   2045-2322
  • 通讯作者地址:   Chinese Acad Fishery Sci
  • 被引频次:   30
  • DOI:   10.1038/s41598-017-05352-7
  • 出版年:   2017

▎ 摘  要

Gold nanoparticles (AuNP) dotted reduction graphene oxide (RGO-AuNP) is used as a platform for an aptamer biosensor to selectively detect 3,3'4,4'-polychlorinated biphenyls (PCB77). By anchoring aptamers onto the binding sites of RGO-AuNP and making use of the synergy effect of RGO and AuNP, the RGO-AuNP based biosensor exhibits superior analytical performances to AuNP based biosensor in terms of sensitivity and repeatability. The sensitivity of RGO-AuNP based aptamers (RGO-AuNP-Ap) biosensor (226.8 mu A cm(-2)) is nearly two times higher than that of Au based biosensors (AuNP-Ap/Au electrode, 147.2 mu A cm(-2)). The RGO-AuNP-Ap/Au biosensor demonstrated a linear response for PCB77 concentrations between 1 pg L-1 and 10 mu g L-1, with a low limit of detection (LOD) of 0.1 pg L-1. The superb LOD satisfies the exposure thresholds (uncontaminated water < 0.1 ng L-1) set out by International Agency for Research on Cancer (IARC) and the Environmental Protection Agency (EPA). The proposed biosensor can be a powerful tool for rapid, sensitive and selective detection of PCBs on site.