▎ 摘 要
In this study, the application of a novel nanomaterial composite was investigated in two microextraction techniques of solid-phase microextraction and a needle trap device in a variety of sampling conditions. The optimum sampling temperature and relative humidity were 10 degrees C and 20%, respectively, for both techniques with two sorbents of graphene/silica composite and polydimethyl siloxane. The two microextraction techniques with the proposed sorbent showed recoveries of 95.2 and 94.6% after 7 days. For the needle trap device the optimums desorption time and temperature were 3 min at 290 degrees C and for SPME these measures were 1 and 1.5 min at 240-250 degrees C for the graphene/silica composite and polydimethyl siloxane, respectively. The relative standard division obtained in inter-and intra-day comparative studies were 3.3-14.3 and 5.1-25.4, respectively. For four sample the limit of detection was 0.021-0.25 ng/mL, and the limit of quantitation was 0.08-0.75 ng/mL. The results show that the graphene/silica composite is an appropriate extraction media for both techniques. Combining an appropriate sorbent with microextraction techniques, and using these in conjunction with a sensitive analytical instrument can introduce a strong method for sampling and analysis of occupational and environmental pollutants in air.